Introduction
The number of double bond and aromatic conjugations inside a molecule is determined using ultraviolet and visible spectroscopy, sometimes referred to as electronic spectroscopy.
In this process, electrons are moved from the HOMO to the LUMO orbital (HOMO stands for Highest Occupied Molecular Orbital and LUMO for Lowest Unoccupied Molecular Orbital). As conjugation increases, the HOMO-LUMO gap narrows.
A popular analytical method known as UV visible spectroscopy uses the absorption of light in the ultraviolet to visible region to provide details about the electrical structure of molecules. Inorganic chemistry, biology, and analytical chemistry all often employ UV-visible spectroscopy.
A sample is exposed to a variety of light wavelengths in the ultraviolet-visible range during UV-visible spectroscopy. This may affect how some sample components absorb or do not absorb photons of particular energies. The molecular structure and chemical make-up of the sample can be ascertained by the absorption or non-absorption of specific light wavelengths.
Absorption and intensity shifts caused by changes in the electronic structure of a sample or changes in the experimental circumstances are important issues to address in UV visible spectroscopy. This article will go into depth about these two events, including their causes and how they impact the spectrum analysis process.
Absorption Shifts
Absorption shifts are variations in the wavelength of a species' maximum absorbance or the placement of absorption peaks within the UV-visible spectrum induced by changes in the sample's electronic structure. Changes in the electrical structure of the sample might cause absorption shifts.
Absorption shifts refer
to changes in the wavelength of the maximum absorbance of a species or the
location of absorption peaks within the UV-visible spectrum, which are caused
by changes in the electronic structure of the sample. Absorption shifts can
occur due to changes in the electronic structure of the sample. A shift to a
longer wavelength indicates an increase in energy required for excitation,
while a shift to a shorter wavelength indicates a decrease in energy required. The
causes of these shifts are varied and include changes in the oxidation state of
a species, the presence of conjugated systems within the sample, and changes in
coordination number.
A) Oxidation State
One cause of absorption
shifts in UV-visible spectroscopy is the change in oxidation state of the
sample. Oxidation is a process that involves the addition of electrons to a
molecule by others, leading to the conversion of this molecule or an atom to
its higher oxidation state. The reverse reaction is called reduction, which
involves the loss of electrons from a molecule.
In cases where a molecule
undergoes oxidation, the transition energy of the molecule may change, causing
the absorption maxima to shift to a longer wavelength. An example of this is the shifting of the Fe^2+ ion from 507 nm to
535 nm upon oxidation to Fe^3+. The electronic structure of Fe^3+ involves a
single unpaired electron based on the complex ion's d orbitals.
The shifting of the
absorption maxima of Fe^2+ to Fe^3+ is due to the formation of a new molecular
orbital (MO) when electrons are added to the complex ion. In Fe^2+ ions, the
energy of the MO is lower, corresponding to a maximum absorbance at 507 nm.
Upon oxidation, the energy of the MO increases, which shifts the maximum
absorbance to a longer wavelength (535 nm).
b) Conjugation
We know that, E = hc As the conjugation increase transition energy (E) between the orbitals is decrease and therefore wavelength (max) increase. If double bonds (chromophore) present in the molecule are in conjugation, then absorption shift towards longer wavelength. In compound 'A', double bonds are in conjugation therefore 'A' possessing higher wavelength (max) as compared to compound 'B' (non-conjugated derivative).
The set of atoms in a molecular structure that alternately possess single and double bonds is referred to as a conjugated system. Conjugation alters the molecule's electrical structure significantly, changing the absorption of UV and visible light.
There is considerable -conjugation in inorganic dyes such food colouring, azo dyes, and acid-base indicators. As a result, the system interacts with light, producing the distinctive colours that can be seen. As an example, the presence of a conjugated double bond in the system causes azo dyes to absorb at longer wavelengths than their comparable aniline dyes.
c) Coordination Number
A third cause of
absorption shifts in UV-visible spectroscopy is a change in coordination
number. Coordination compounds such as metal complexes can undergo changes in
their coordination number that lead to changes in their UV-visible absorbance.
For example, the
coordination number of Co in a complex containing a diamine ligand is four. Its
electron configuration is d6, and this requires ligands with low-energy
orbitals, such as halides. However, when the number of ligands increases, the
energy required to excite an electron from the t2g set to the eg set decreases.
The energy required for the d-d transition is lowered upon the formation of the
hexacoordinate complex, leading to a maximum absorbance at a longer wavelength
(from 515 to 550 nm).
Intensity Shifts
Intensity shifts refer to
changes in the intensity of the absorbance peaks. Intensity shifts are possible
and can indicate changes in the sample or experimental conditions. An increase
in intensity may be due to changes in the concentration of the analyte or the
introduction of a co-solvent. In contrast, a decrease in intensity may be
indicative of sample degradation, decreased analyte concentration, or the
presence of a scavenging agent.
a) Concentration
A change in concentration
can lead to an intensity shift in the UV-visible spectrum. When a sample's
concentration is too low, the signals may be weak or barely detectable. On the
other hand, high concentrations of a sample can lead to overloading of the
detector or produce signal saturation. Both effects reduce the sensitivity of
the analysis and can lead to erroneous results.
The sensitivity of the
analysis can be controlled by changing the sample concentration, but it is
challenging to find the optimum concentration. The optimum concentration is
usually dependent on the instrument's sensitivity, the wavelength range, and
the sample's identity.
b) Environment
The sample environment
can also play a role in intensity shifts in UV-visible spectroscopy. The
environment may include the type of solvent used, temperature, or pressure.
These factors can lead to changes in the sample's structure or chemical
configuration, leading to a change in the sample's spectral behavior.
For example, an increase
in temperature may cause a shift in the maximum absorbance wavelength due to a
change in the electronic structure of the sample. The pH of the solvent used
can also affect the maximum absorbance, depending on the pKa of the sample's
constituents.
c) Aggregation
Aggregation can also
cause intensity shifts in UV-visible spectroscopy. Aggregation occurs when
molecules in solution come together to form larger particles. The formation of
aggregates can cause a shift in the absorbance spectrum since the interaction
between the particles can affect the electronic structure of the sample.
For example, aggregation
of proteins occurs when proteins come together to form multimers. This can lead
to changes in the maximum absorbance of the protein due to the changes in the
electronic structure of the protein as the proteins aggregate.
Conclusion
In conclusion, UV-visible
spectroscopy is a powerful analytical technique used in the identification and
quantification of analytes in a sample. Spectral changes in absorption and
intensity of UV-visible light can be used to identify and quantify the analytes
concentration present in the sample. Shifts in the absorbance maxima and
intensity of UV-visible spectra can be caused by changes in the electronic
structure of the sample, solution environment, and sample concentration.
Analysis of these shifts can be crucial in providing insights into the
molecular structure and composition of the sample under investigation.
No comments:
Post a Comment